From the line to a result with its uncertainty
The least-squares slope b and intercept a minimise the squared residuals of the signals. The unknown’s concentration is x₀ = (ȳ₀ − a)/b, where ȳ₀ is the mean of m replicate readings. Its standard error grows with the scatter about the line (s_y/x), falls with more replicates and more standards, and is smallest near the centre of the curve, which is why unknowns should fall in the middle of the standards rather than at the ends. The confidence interval is ± t × s(x₀) with n − 2 degrees of freedom. A high R² alone does not prove linearity: look for a pattern in the residuals.
Detection and quantitation limits
The limit of detection (LOD) and limit of quantitation (LOQ) here use the ICH approach based on the calibration line: 3.3 and 10 times the residual standard deviation, divided by the slope. They are estimates for the measured solution; confirm them with replicate blanks or low standards when a method is validated.
Assumptions and limits
- The response is linear over the range of the standards and the scatter is the same at all concentrations (unweighted least squares).
- Errors are in the signal, not in the standard concentrations.
- Replicate readings of the unknown are averaged; their spread is not used in the error estimate.
- Through-origin fits use n − 1 degrees of freedom and a first-order error estimate.
References
- Miller, J. N., Miller, J. C. and Miller, R. D. (2018), Statistics and Chemometrics for Analytical Chemistry, 7th ed., Pearson, chapter 5 (calibration methods in instrumental analysis: regression and correlation).
- Harris, D. C. and Lucy, C. A. (2020), Quantitative Chemical Analysis, 10th ed., W. H. Freeman, chapter on the method of least squares and calibration curves.
- International Council for Harmonisation (2005), ICH Q2(R1) Validation of Analytical Procedures: Text and Methodology (detection and quantitation limits based on the standard deviation of the response and the slope).
- Abramowitz, M. and Stegun, I. A. (1964), Handbook of Mathematical Functions, National Bureau of Standards, 26.7 (Student’s t-distribution).
Questions
- What concentration gives a reading of 0.263 on the example curve?
- With standards 0 to 10 and readings 0.262, 0.268 and 0.259, the line is y = 0.003238 + 0.05169x, so the unknown is 5.026 ± 0.072 (95 %). R² is 0.99992 and the limit of quantitation is 0.367.
- Should I force the calibration line through zero?
- Only if the blank really gives zero signal and the intercept is not significantly different from zero. Forcing it otherwise biases low concentrations; the default line with an intercept is safer.
- What R² is good enough for a calibration curve?
- Many methods ask for R² of at least 0.99 or 0.995, but check the residuals too: a curve can have a high R² and still bend.
This calculator is a design aid. Check results against the governing code or standard and have critical work verified by a qualified engineer. Read the disclaimer.