Centrifuge RCF to RPM Calculator (g-Force)

Protocols give centrifugation in × g because the force depends on both speed and rotor radius, while centrifuges are set in rpm. This calculator converts either way for your rotor, and gives the k-factor and an estimated pelleting time for particles of known sedimentation coefficient.

Inputs

Advanced options (2)

Results

Result
1,006 × g
RCF at r_max
1,006.4
× g
Rotor speed
3,000
Linear speed at the radius
31.42
k-factor (clearing factor)
19,485
Pelleting time t = k/S
1,169

    RefBIO-CEL-003
    MethodRCF = ω²r/g = 1.11824 × 10⁻⁵ × r(cm) × rpm²
    UnitsSI
    Date2026-11-24

    Working

    Calculated in SI base units; inputs are converted from the units you choose.

    1. RCF from speedRCF = ω²r/g, ω = 2π·rpm/60RCF = (314.2 rad/s)² × 0.1 m / 9.80665 = 1,006 × g
    2. k-factork = 2.53 × 10¹¹ × ln(r_max/r_min)/rpm²; t (h) = k/Sk = 19,485, t = 19.49 h for S = 1,000

    Radius matters

    RCF = ω²r/g, so at the same speed a larger rotor gives a higher force, growing with the square of the speed. Use the maximum radius, at the bottom of the tube, which is printed on most rotors. The k-factor describes how efficiently a rotor pellets particles: the time in hours to pellet particles with sedimentation coefficient S (in svedbergs) is about k/S.

    Assumptions and limits

    • Radius measured from the centre of rotation to the bottom of the tube (r_max).
    • Pelleting time assumes particles in water at 20 °C with the stated S value.
    • Never exceed the rated speed of the rotor or tubes.

    References

    • Beckman Coulter, Rotors and Tubes for Preparative Ultracentrifuges (k-factor and RCF definitions).
    • Rickwood, D. and Graham, J. M., Biological Centrifugation, BIOS Scientific Publishers, 2001.

    Questions

    What RCF does 3000 rpm give in a 10 cm rotor?
    1006 × g.
    What speed gives 1000 × g in a 10 cm rotor?
    2990 rpm.

    This calculator is a design aid. Check results against the governing code or standard and have critical work verified by a qualified engineer. Read the disclaimer.