Why multiply by 10,000
Each large square of an improved Neubauer hemocytometer is 1 mm × 1 mm with a 0.1 mm deep chamber, holding 0.1 µL or 10⁻⁴ mL. The average count per square, times 10⁴, is cells per mL in the chamber; multiplying by the dilution (2 for an equal volume of trypan blue) gives the density of the original suspension. Live cells exclude the dye and stay clear; dead cells take it up and look blue.
Assumptions and limits
- Improved Neubauer chamber with 0.1 µL large squares.
- Cells evenly suspended and not clumped; count cells touching two chosen borders only.
- Trypan blue counts read within a few minutes, before live cells start taking up the dye.
References
- Freshney, R. I., Culture of Animal Cells: A Manual of Basic Technique and Specialized Applications, 7th ed., Wiley-Blackwell, 2016, chapter on quantitation.
- Strober, W. (2015), Trypan blue exclusion test of cell viability, Current Protocols in Immunology 111, A3.B.1–A3.B.3.
Questions
- What is the density if I count 200 live and 20 dead cells in 4 squares after a 1:1 trypan blue dilution?
- 1.0 × 10⁶ viable cells/mL with 90.9 % viability, so 10 mL of suspension holds 1.0 × 10⁷ live cells.
- How do I seed 10 mL at 5 × 10⁴ cells/mL from that suspension?
- Take 0.5 mL of the suspension and add 9.5 mL of medium.
This calculator is a design aid. Check results against the governing code or standard and have critical work verified by a qualified engineer. Read the disclaimer.